產(chǎn)品規(guī)格參數(shù)
-
Measured Variables
Concentration
Dew Point
Cell Pressure
Cell Temperature -
Hazardous area approvals
ATEX / IECEx Zone 1
UKEx Zone 1
CSA Class I, Division 1
CSA Class I, Zone 1
Measured Variables
Concentration
Dew Point
Cell Pressure
Cell Temperature
Hazardous area approvals
ATEX / IECEx Zone 1
UKEx Zone 1
CSA Class I, Division 1
CSA Class I, Zone 1
-
測量原理
TDLAS
-
Product Headline
Best-in-class gas analyzer for moisture (H2O) that is exceptionally reliable and tailored for the natural gas industry. The sensor measures gas using a Tunable Laser Diode Absorption Spectroscopy (TDLAS) to determine the concentration of the gas without coming into physical contact with the stream. Heartbeat Technology ensures compliance and process safety at all times.
-
Channels
1
-
Analyte and Measurement ranges
H2O (Moisture): 0-50 to 0-6000 ppmv
-
Measured Variables
Concentration
Dew Point
Cell Pressure
Cell Temperature -
Ambient Temperature range
-20 to 60°C (-4 to 140°F)
-
Operating Pressure range
Inlet Pressure: 140-310 kPa (20-45 psig)
Sample Cell: 800-1200 mbara or 800-1700 mbara (optional) -
Analyzer Wetted materials
316L Stainless Steel
FKM O-Rings
Glass -
Power supply
Without heater: 24 VDC ±20% or 100 to 240 VAC ±10%, 50/60 Hz, 10W
With optional heater: 100 to 240 VAC ±10%, 50/60 Hz, 80W -
Communication
Service webserver interface: Ethernet RJ45
I/O1: Modbus RTU over RS485
I/O2 and 3: Relay output OR Universal I/O (UIO); UIO can be configured as analog output (4-20 mA) or digital/status output -
Housing materials
Electronics: Copper-free aluminum
Sample System Enclosure: 304 Stainless Steel
Sample System Panel: Anodized aluminum -
Hazardous area approvals
ATEX / IECEx Zone 1
UKEx Zone 1
CSA Class I, Division 1
CSA Class I, Zone 1 -
Degree of protection
IP66, Type 4X
-
Product safety
CE
RCM
FCC